FDA Adverse Event
Malfunction
Summary report: N
ENDOCARE CRYOCARE CRYOPROBE 1.7MM RIGHT ANGLE
MDR report key: 2933875
·
Received January 29, 2013
Report
- Report Number
- 3008262715-2013-00002
- Event Type
- Malfunction
- Date Received
- January 29, 2013
- Date of Event
- December 31, 2012
- Report Date
- January 2, 2013
- Manufacturer
- HEALTHTRONICS, INC.
- Product Code
- GEH
- PMA / PMN Number
- K011074
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- RADIOLOGIC TECHNOLOGIST
Narratives
Additional Manufacturer Narrative · 1
NO PATIENT INJURY REPORTED. PROBE RETURNED TO MANUFACTURER AND EVALUATED. REPORTED DISCREPANCY COULD NOT BE DUPLICATED. RETURNED PROBE MET ALL MANUFACTURING SPECIFICATIONS.SERVICE TECHNICIAN SENT OUT TO HOSPITAL TO CHECK THE CRYO-CS UNIT (CSL 1024). COULD NOT GET SPECIFICS ON ORIGINAL PROBLEM. COMPLETED ANNUAL PM, UNIT OPERATING WITHIN SPECIFICATION.
Description of Event or Problem · 1
AT THE TIME OF FREEZING, COULD NOT MAINTAIN PROPER TEMPERATURE - COMPLETE SHUTDOWN AND MULTIPLE PROBE CONNECTION CHANGES NEEDED. THE PROBE KEPT DISENGAGING AND AFTER EACH FUNCTION CHANGE, THE DOCTOR WOULD NEED TO CHANGE PORTS TO GET THE PROBE TO FUNCTION PROPERLY, 60 MINUTE DELAY IN CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 38547 | ENDOCARE CRYOCARE CRYOPROBE 1.7MM RIGHT ANGLE | UNIT, CRYOSURGICAL, ACCESSORIES | GEH | HEALTHTRONICS, INC. | PERC-17 | 9602 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |