FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 2933846 · Received January 29, 2013

Report

Report Number
2032227-2013-00382
Event Type
Malfunction
Date Received
January 29, 2013
Date of Event
January 14, 2013
Report Date
January 14, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP HAD NORMAL OPERATING CURRENTS AND NO OFF NO POWER, LOW BATTERY OR BATTERY OUT LIMIT ALARMS NOTED. THE INSULIN PUMP HAD MISSING END CAP STICKER.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A FROZEN DISPLAY AND UNRESPONSIVE BUTTONS. THE CUSTOMER STATED THAT THE INSULIN PUMP HAD THE ALARM BATTERY OUT LIMIT ON THE SCREEN BUT SHE WAS UNABLE TO CLEAR IT. TROUBLESHOOTING WAS PERFORMED, BUT THE ISSUES WERE NOT RESOLVED. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
39025 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAS

Patients

Seq Age Sex Outcome Treatment
1 33 YR