FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2933807 · Received January 29, 2013

Report

Report Number
2531779-2013-01326
Event Type
Malfunction
Date Received
January 29, 2013
Report Date
January 18, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 05/06/2015 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR DAMAGE WAS OBSERVED. DURING TESTING, ALL OF THE KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY UNRESPONSIVE. THE KEYPAD COVER WAS REMOVED AND THERE WAS EVIDENCE OF CONTAMINATION FOUND UNDER ALL OF THE KEY CONTACTS. UNRELATED TO THE COMPLAINT, INVESTIGATION REVEALED A DIM DISPLAY WITH DISCOLORED TEXT AND MULTIPLE CRACKS IN THE BATTERY COMPARTMENT. ALSO UNRELATED TO THE COMPLAINT, THE AUDIO BOLUS BUTTON WAS FOUND TO BE UNRESPONSIVE. THERE WAS NO DAMAGE OBSERVED IN THE AUDIO BOLUS BUTTON.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(4) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER STATED THAT THE UP ARROW, DOWN ARROW AND OK KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE AND REQUIRED MULTIPLE BUTTON PRESSES TO ELICIT A RESPONSE. IT WAS NOTED THAT THE OK BUTTON WAS HYPER-RESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
39557 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 57 YR