FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 2933795 · Received January 29, 2013

Report

Report Number
1416980-2013-02196
Event Type
Malfunction
Date Received
January 29, 2013
Date of Event
December 10, 2012
Report Date
January 16, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS ONE OR MORE CYCLE ADVANCED TO NEXT FILL WHEN SLOW/NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING THE EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED DURING THE THERAPY INITIATED ON (B)(6) 2012, AT 21:06:48. DURING NIGHT DRAIN CYCLE THREE, THE PATIENT'S ULTRAFILTRATION READING WAS 1559ML, INDICATING THE HOME PATIENT (HP) DRAINED 1559ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
39610 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1