FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2933786 · Received January 29, 2013

Report

Report Number
1416980-2013-02193
Event Type
Malfunction
Date Received
January 29, 2013
Date of Event
December 13, 2012
Report Date
December 13, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS REPAIRED ON-SITE AT THE CUSTOMER FACILITY BY A BAXTER FIELD SERVICE TECHNICIAN, AND THE CONDITION WAS CONFIRMED. THIS IS AN ANCILLARY SERVICE EVENT DISCOVERED DURING PREVENTATIVE MAINTENANCE. THE CAUSE WAS DETERMINED TO BE BATTERY DISCHARGE BELOW THE THRESHOLD. TO CORRECT THE CONDITION, THE MAIN BATTERY WAS REPLACED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE A DAMAGED BATTERY. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THIS WAS NOT REPORTED BY THE CUSTOMER. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
38927 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1