ARCHITECT CA 19-9XR
Report
- Report Number
- 1415939-2013-00026
- Event Type
- Malfunction
- Date Received
- January 29, 2013
- Date of Event
- January 21, 2013
- Report Date
- January 21, 2013
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- NIG
- PMA / PMN Number
- K052000
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THE TREND REVIEW IDENTIFIED NORMAL COMPLAINT ACTIVITY FOR THE ISSUE UNDER INVESTIGATION. ACCURACY TESTING WAS COMPLETED TO EVALUATE THE ASSAY PERFORMANCE OF LOT 18067M500. THE TESTING MET ACCEPTANCE CRITERIA. THE INVESTIGATION RESULTS SHOW THAT THERE IS NO PRODUCT DEFICIENCY AND THAT THE ARCHITECT CA 19-9XR REAGENTS ARE PERFORMING AS INTENDED. FURTHER INVESTIGATION IS NOT REQUIRED. NO MALFUNCTION WAS IDENTIFIED. THIS ISSUE WAS LIMITED TO A DISCREET PATIENT SAMPLE. RETEST OF THE SAMPLE WITH THE SAME LOT NUMBER GENERATED AN ACCURATE VALUE. ACCURACY TESTING WAS COMPLETED USING PANELS AND THE LIKELY CAUSE LOT SHOWS THAT IT CAN ACCURATELY DETECT KNOWN CONCENTRATIONS OF THE ANALYTE.
(B)(4). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. COMPLETE SAMPLE IDENTIFICATION NUMBER IS (B)(6). THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 02K91-25 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 02K91-27.
THE CUSTOMER STATED THAT THE PATIENT SAMPLE GENERATED A FALSELY ELEVATED ARCHITECT CA19-9 ASSAY RESULT OF 165 U/ML. THE CUSTOMER PERFORMED MULTIPLE RUNS ON THE SAMPLE WITH THE SAME REAGENT LOT, GENERATING THE FOLLOWING RESULTS: REPEAT RESULTS (U/ML) LESS THAN 22.833. CUSTOMER REPORTED FROM THE LAB THE VALUE OF LESS THAN 2. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED. NO ADDITIONAL PATIENT INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 38970 | ARCHITECT CA 19-9XR | NIG | ABBOTT LABORATORIES | 18067M500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ARCHITECT I2000SR, LN 03M74-01| SN (B)(4)| SN (B)(4)| ARCHITECT I2000SR, LN 03M74-01 |