FDA Adverse Event Malfunction Summary report: N

SECURE II MED/SURG BED

MDR report key: 2933467 · Received December 21, 2012

Report

Report Number
1831750-2012-13151
Event Type
Malfunction
Date Received
December 21, 2012
Date of Event
November 29, 2012
Report Date
November 29, 2012
Manufacturer
STRYKER MEDICAL
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

MOTION INTERRUPT PAN BOLTED TOO TIGHT. CPU BOARD NOT BURNED-IN AFTER REPLACEMENT BY CUSTOMER. SEE SCANNED PAGE.

Description of Event or Problem · 1

IT WAS REPORTED BY SERVICE REPORT THAT THE MOTION INTERRUPT PAN AND BED LIFT WERE STUCK AND THE BED EXIT WAS NOT FUNCTIONING. NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES ARE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SECURE II MED/SURG BED A/C POWERED HOSPITAL BED FNL STRYKER MEDICAL 3002 EX NA

Patients

Seq Age Sex Outcome Treatment
1