FDA Adverse Event Injury Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2933447 · Received January 29, 2013

Report

Report Number
2531779-2013-01303
Event Type
Injury
Date Received
January 29, 2013
Date of Event
December 29, 2012
Report Date
January 5, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4):A REVIEW OF THE TOTAL DAILY DOSE HISTORY INDICATED THAT INSULIN DELIVERY TOTALS CORRECTLY REFLECTED PROGRAMMED VALUES. THE PUMP WAS EXERCISED FOR 29 HOURS WITH NO DELIVERY ISSUES. THE PUMP WAS EVALUATED AND FOUND TO BE OPERATING WITHIN REQUIRED SPECIFICATIONS AND DELIVERING INSULIN ACCURATELY.

Description of Event or Problem · 1

THE REPORTER CONTACTED ANIMAS ON (B)(6) 2013 STATING THAT THE PATIENT WAS HOSPITALIZED FOR A BLOOD GLUCOSE OF 479 MG/DL WITH HIGH KETONES AND VOMITING. THE REPORTER INDICATED THAT THE PATIENT WAS RESPONDING TO INJECTIONS BUT BLOOD GLUCOSE WOULD GO BACK UP WHEN ON THE PUMP. THE REPORTER CONFIRMED NO KNOWN ISSUES WITH THE SITE OR SET. THE PUMP WAS REVIEWED WITH THE PATIENT. THE PATIENT CONFIRMED NO RELEVANT ALARMS, THE BOLUS HISTORY WAS CORRECT, TOTAL DAILY DOSE AMOUNTS WERE CORRECT, THE SUSPEND HISTORY WAS CORRECT, AND THE BASAL HISTORY WAS CORRECT. THE REPORTER STATED THAT THEY HAD LOST FAITH IN THE PUMP. THIS REPORT IS MADE BASED ON THE ALLEGATION THAT THE PATIENT WAS HOSPITALIZED FOR HYPERGLYCEMIA RELATED TO A POSSIBLE ISSUE WITH THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
39228 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 16 YR Hospitalization| L