ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-01290
- Event Type
- Malfunction
- Date Received
- January 29, 2013
- Report Date
- January 7, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #2: DATE OF SUBMISSION 11/27/2013. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/19/2013 WITH THE FOLLOWING FINDINGS: TESTING CONFIRMED INTERMITTENT RESPONSES TO BUTTON PRESSES ON THE UP AND DOWN KEYPAD BUTTONS. MULTIPLE BUTTON PRESSES WERE REQUIRED BEFORE THE UP AND DOWN BUTTONS ENGAGED. THE UP BUTTON WAS HARD TO PRESS AND REQUIRED INCREASED FORCE TO ENGAGE. THE OK AND CONTRAST BUTTONS RESPONDED NORMALLY TO USER INPUT. THERE WAS NO VISIBLE DAMAGE ON THE KEYPAD. CONTAMINATION WAS PRESENT UNDER THE CONTACTS OF ALL BUTTONS. UNRELATED TO THE INITIAL COMPLAINT, THE PUMP BOOTED TO THE VERIFY SCREEN WITH A DIM PINK CONTRAST. THE OLED SCREEN WAS REMOVED AND REPLACED WITH A NEW TEST SCREEN WHICH CAUSED THE CONTRAST TO RETURN TO NORMAL. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
(B)(4).
ON (B)(4) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 39018 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 3 YR |