FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2933261 · Received January 29, 2013

Report

Report Number
1416980-2013-02153
Event Type
Malfunction
Date Received
January 29, 2013
Date of Event
December 6, 2012
Report Date
December 6, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: EVALUATION SUMMARY: THE REPORTED CONDITION OF "F-38 ALARM" WAS CONFIRMED DURING DEVICE EVALUATION. THE ROOT CAUSE WAS DETERMINED TO BE DUE TO DEFECTIVE TUBE MISLOADING SENSORS. THE TUBE MISLOADING SENSORS WERE REPLACED TO CORRECT THE REPORTED CONDITION.

Additional Manufacturer Narrative · 1

(B)(4)-(CORRECTION TO INITIAL MDR) IT IS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT; HOWEVER, THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. THIS DEVICE WAS EVALUATED ON-SITE BY A BAXTER FIELD SERVICE TECHNICIAN. UPON COMPLETION OF BAXTER'S INVESTIGATION, A FOLLOW-UP WILL BE SUBMITTED. A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS CUSTOMER SERVICE REQUEST ON THIS PUMP.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER MEXICO THAT A FLOGARD INFUSION PUMP HAD AN F-38 ALARM. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT; THEREFORE, NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.