FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 2933171 · Received January 29, 2013

Report

Report Number
0001811755-2013-00088
Event Type
Malfunction
Date Received
January 29, 2013
Date of Event
January 11, 2013
Report Date
January 11, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING THE DEVICE EVALUATION, THE HYBRID BALL BEARING WAS FOUND TO BE CORRODED. INCREASED FRICTION FROM THE CORROSION IS A PROBABLE CAUSE OF THE REPORTED OVERHEATING OF THE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY, THE DRILL WAS OVERHEATING. BACK-UP EQUIPMENT WAS USED TO COMPLETE THE PROCEDURE. NO DELAY, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION WERE REPORTED.