HEARTSINE SAMARITAN 300P AND PAD PAK
Report
- Report Number
- 3004123209-2013-00150
- Event Type
- Malfunction
- Date Received
- January 23, 2013
- Date of Event
- December 5, 2010
- Report Date
- January 14, 2013
- Manufacturer
- HEARTSINE TECHNOLOGIES, LTD
- Product Code
- MKJ
- PMA / PMN Number
- K052465
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EI
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE PAD DEVICE WAS INSTALLED ON (B)(6) 2008 AND OPERATED SUCCESSFULLY UNTIL A FAILED SELF TEST ON (B)(6) 2010 DUE TO A LOW BATTERY. THE TEMPERATURE RECORDED AROUND THIS TIME WAS AS LOW AS 2 DEGREES C INDICATING THE DEVICE WAS NOT BEING ADEQUATELY STORED. THERE IS THEN A GAP IN THE HISTORY UNTIL (B)(6) 2011, INDICATING THE DEVICE WAS LEFT IN FAULT MODE OR THE PAD-PAK WAS REMOVED. AFTER THIS DATE THERE ARE MULTIPLE EVENTS ON THE SAME DAY WHICH WOULD INDICATE THE DEVICE WAS SWITCHING ITSELF ON AUTOMATICALLY. THE PROBLEM WITH THE DEVICE WAS ATTRIBUTED TO A FAULT IN THE MEMBRANE. THE LOW BATTERY WARNINGS WERE ISSUED BECAUSE THE PAD-PAK WAS DEPLETED TO THE POINT IT TRIGGERED THE BATTERY MANAGEMENT SYSTEM IN THE DEVICE. THE BATTERIES CAN BE DEPLETED BY A VARIETY OF CONDITIONS INCLUDING THE STORAGE LOCATION OF THE DEVICE, AGE OF BATTERY PACK AND LEVEL OF MANUAL INTERVENTION. THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.
THERE WAS NO PATIENT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE IT WAS SWITCHING ITSELF ON AUTOMATICALLY. A DEVICE SWITCHING ITSELF ON AUTOMATICALLY, IF LEFT UNDETECTED COULD RESULT IN THE BATTERY BECOMING DEPLETED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 33394 | HEARTSINE SAMARITAN 300P AND PAD PAK | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES, LTD | PAD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |