FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2932790 · Received January 29, 2013

Report

Report Number
2531779-2013-01281
Event Type
Malfunction
Date Received
January 29, 2013
Report Date
January 4, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013, WITH THE FOLLOWING FINDINGS: DURING THE LOAD CARTRIDGE STEP, THE PUMP PUSHED OUT HALF OF THE CARTRIDGE PRIOR TO STOPPING. THE PUMP WAS OPENED AND CONTAMINATION WAS FOUND WITHIN THE FORCE SENSOR ASSEMBLY. UNRELATED TO THE COMPLAINT, EVIDENCE OF MOISTURE DAMAGE WAS FOUND ON THE PUMP MOTOR CIRCUIT; THE MOTOR WAS TESTED AND NO DEFECTS WERE FOUND. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED TO ANIMAS. INVESTIGATION FOUND CONTAMINATION IN THE FORCE SENSOR ASSEMBLY. THIS REPORT IS MADE BASED ON THE RESULTS OF INVESTIGATION.