FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2932790
·
Received January 29, 2013
Report
- Report Number
- 2531779-2013-01281
- Event Type
- Malfunction
- Date Received
- January 29, 2013
- Report Date
- January 4, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013, WITH THE FOLLOWING FINDINGS: DURING THE LOAD CARTRIDGE STEP, THE PUMP PUSHED OUT HALF OF THE CARTRIDGE PRIOR TO STOPPING. THE PUMP WAS OPENED AND CONTAMINATION WAS FOUND WITHIN THE FORCE SENSOR ASSEMBLY. UNRELATED TO THE COMPLAINT, EVIDENCE OF MOISTURE DAMAGE WAS FOUND ON THE PUMP MOTOR CIRCUIT; THE MOTOR WAS TESTED AND NO DEFECTS WERE FOUND. (B)(6).
Description of Event or Problem · 1
THE PUMP WAS RETURNED TO ANIMAS. INVESTIGATION FOUND CONTAMINATION IN THE FORCE SENSOR ASSEMBLY. THIS REPORT IS MADE BASED ON THE RESULTS OF INVESTIGATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 38859 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 2 YR |