FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 2932771 · Received January 23, 2013

Report

Report Number
3004123209-2013-00191
Event Type
Malfunction
Date Received
January 23, 2013
Date of Event
July 3, 2011
Report Date
January 16, 2013
Manufacturer
HEARTSINE TECHNOLOGIES, LTD
Product Code
MKJ
PMA / PMN Number
K052465
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RP
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PAD DEVICE WAS INSTALLED ON (B)(6) 2007 AND OPERATED SUCCESSFULLY UNTIL A FAILED SELF TEST ON (B)(6) 2011 DUE TO A LOW BATTERY. THE DEVICE FAILED ITS WEEKLY SELF A FURTHER 4 TIMES BETWEEN (B)(6) 2011, ALL FOR LOW BATTERIES. AFTER THIS DATE, THERE ARE MULTIPLE EVENTS ON THE SAME DAY, WHICH WOULD INDICATE THE DEVICE WAS SWITCHING ITSELF ON AUTOMATICALLY. NO FAULT WAS FOUND WITH THE RETURNED PAD DEVICE. THE PAD-PAK USED IN THE DEVICE WAS NOT RETURNED FOR INVESTIGATION. THE PAD-PAK WAS LIKELY TO BE CLOSE, IF NOT PAST, ITS EXPIRY DATE. THE LOW BATTERY WARNINGS WERE DELIVERED BECAUSE THE PAD-PAK WAS DEPLETED TO THE POINT IT TRIGGERED THE BATTERY MANAGEMENT SYSTEM IN THE DEVICE. THE BATTERIES CAN BE DEPLETED BY A VARIETY OF CONDITIONS INCLUDING THE STORAGE LOCATION OF THE DEVICE, AGE OF BATTERY PACK, AND LEVEL OF MANUAL INTERVENTION. THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELED FOR SINGLE USE, BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.

Description of Event or Problem · 1

THERE WAS NO PT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE IT WAS SWITCHING ITSELF ON AUTOMATICALLY. A DEVICE SWITCHING ITSELF ON AUTOMATICALLY, IF LEFT UNDETECTED, COULD RESULT IN THE BATTERY BECOMING DEPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
33096 HEARTSINE SAMARITAN 300P AND PAD PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES, LTD PAD

Patients

Seq Age Sex Outcome Treatment
1