FDA Adverse Event Malfunction Summary report: N

ENDOPATH** XCEL* STABILITY SLEEVE

MDR report key: 2932718 · Received January 29, 2013

Report

Report Number
3005075853-2013-00355
Event Type
Malfunction
Date Received
January 29, 2013
Report Date
January 11, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GCJ
PMA / PMN Number
K032676
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. ADDITIONAL INFORMATION: DOES THIS TROCAR HAVE THE OPTIVIEW TECHNOLOGY? ASKU. WERE ANY NOISES HEARD SUCH AS "WHISTLING" OR "HISSING"? ASKU. IF SO, DID THE "NOISE" PREVENT INSUFFLATION? ASKU. WAS THERE A DROP IN PRESSURE? ASKU. WHAT WAS THE GAS CONSUMPTION RATE OR VOLUME (LITER/MINUTE)? ASKU. WERE YOU ABLE TO IDENTIFY WHERE THE LEAK WAS COMING FROM? ASKU. WAS A DEVICE INSERTED IN THE TROCAR DURING THE LEAKING? ASKU. WAS ANY TORQUE BEING APPLIED TO THE TROCAR OR DEVICE? ASKU.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS FUNCTIONALLY TESTED TO DETECT ANY LEAKING ISSUES. A LEAK TEST WAS PERFORMED AND THE DEVICE WAS NOTED TO LEAK DURING INSERTION AND REMOVAL OF THE TEST PROBE THROUGH THE DEVICE. THE DEVICE DOES NOT HAVE AN OBTURATOR. THE OBTURATOR IS THE PIECE OF THE TROCAR THAT HAS THE BATCH STAMPED. NO BATCH NUMBER WAS AVAILABLE; THEREFORE, WE WERE UNABLE TO CHECK MANUFACTURING RECORDS FOR ANY RELATED NON-CONFORMANCE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE TROCAR WAS LEAKING. IT IS UNKNOWN IF THERE WAS A DEVICE INSERTED WHEN THE LEAKING OCCURRED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE WITH NO PATIENT CONSEQUENCE.