FDA Adverse Event Malfunction Summary report: N

CERVIOS CHRONOS CURVED SIZ 7

MDR report key: 2932669 · Received January 29, 2013

Report

Report Number
8030965-2013-00242
Event Type
Malfunction
Date Received
January 29, 2013
Report Date
January 4, 2013
Manufacturer
SYNTHES GMBH
Product Code
KWP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES, BUT IS SIMILAR TO DEVICE MARKETED IN THE USA. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS THE PRODUCT IS ENTERING THE COMPLAINT SYSTEM. THE MANUFACTURING DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND.

Additional Manufacturer Narrative · 1

INVESTIGATION COORDINATED BY SYNTHES (B)(4). REPORT RECEIVED INDICATES TESTS PERFORMED DID NOT REVEAL ANY PROBLEMS WITH THE CHRONOS IMPLANT AS WELL AS ALL THE TRIAL IMPLANTS. ALL ARTICLES HAVE BEEN MEASURED, CHECKED AND FOUND TO BE ACCORDING TO THE GIVEN SPECIFICATION. IT IS NOTED THAT THE TRIAL IMPLANTS HAVE A FLAT SURFACE NOT LIKE THE CHRONO IMPLANT ITSELF WHERE WE HAVE TEETH. THEREFORE, IT COULD GET THE IMPRESSION, AFTER THE TEETH SUBSIDED, THAT THE IMPLANT IS LOOSE AND DOES NOT FIT ACCORDINGLY. THE RETURNED ARTICLES WERE ANALYZED FOR CONFORMANCE TO PRINT SPECIFICATION, AS WELL AS THE DEVICE HISTORY RECORDS WERE RESEARCHED. NO ABNORMAL FINDINGS WERE IDENTIFIED. NO PRODUCT FAULT COULD BE DETECTED.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: THE CERVIOS CHRONOS CURVED IMPLANT SIZE 7MM SEEMS TO BE SMALLER THAN TRIAL CERVIOS TRIAL IMPLANT CURVED SIZE 7 AND CORRESPONDS TO CERVIOS TRIAL IMPLANT CURVED SIZE 6. THE IMPLANT IS LABELS 7MM ON THE BOX AS WELL AS THE IMPLANT. THE TRIAL IMPLANTS WERE RECEIVED INTACT. THIS IS 1 OF 5 REPORTS FOR THE SAME EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
39762 CERVIOS CHRONOS CURVED SIZ 7 CERVIOS CHRONOS CURVED KWP SYNTHES GMBH 2640502

Patients

Seq Age Sex Outcome Treatment
1 CERVIOS TRIAL IMPLANTS