HEARTSINE SAMARITAN 300P AND PAD PAK
Report
- Report Number
- 3004123209-2013-00041
- Event Type
- Malfunction
- Date Received
- January 23, 2013
- Date of Event
- May 15, 2011
- Report Date
- January 4, 2013
- Manufacturer
- HEARTSINE TECHNOLOGIES, LTD, CANBERRA HOUSE
- Product Code
- MKJ
- PMA / PMN Number
- K052465
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE PAD DEVICE WAS INSTALLED ON (B)(6) 2010 AND OPERATED SUCCESSFULLY UNTIL (B)(6) 2011. AFTER THIS DATE THERE ARE MULTIPLE EVENTS ON THE SAME DAY WHICH WOULD INDICATE THE DEVICE WAS SWITCHING ITSELF ON AUTOMATICALLY. THE PROBLEM WITH THE DEVICE WAS ATTRIBUTED TO A FAULT IN THE MEMBRANE. A CLOSER INSPECTION OF THE DEVICE HISTORY LOG REVEALED THE TEMPERATURE AT THE TIME OF THE WEEKLY SELF TESTS WAS AS LOW AS -6.4 DEGREE C. EXPOSURE OF THE DEVICE TO ADVERSE ENVIRONMENT CONDITIONS CAN HAVE DETRIMENTAL EFFECTS ON THE MEMBRANE. THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.
THERE WAS NO PT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE IT WAS SWITCHING ITSELF ON AUTOMATICALLY. A DEVICE SWITCHING ITSELF ON AUTOMATICALLY, IF LEFT UNDETECTED COULD RESULT IN THE BATTERY BECOMING DEPLETED.