FDA Adverse Event Malfunction Summary report: N

LIBERTY CYCLER SET, DUAL PATIENT CONENCT

MDR report key: 2932274 · Received January 22, 2013

Report

Report Number
8030665-2013-00052
Event Type
Malfunction
Date Received
January 22, 2013
Date of Event
December 25, 2012
Report Date
December 26, 2012
Manufacturer
REYNOSA MANUFACTURING
Product Code
FKX
PMA / PMN Number
K043363
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION AND THE FAILURE MODE CANNOT BE CONFIRMED. HOWEVER, AN INVESTIGATION OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED BY THE MANUFACTURER. THERE WERE NO DEVIATIONS OR NONCONFORMANCES DURING THE MANUFACTURING PROCESS. IN ADDITION, THE BATCH RECORD REVIEW CONFIRMED THE LABELING, MATERIAL, AND PROCESS CONTROLS WERE WITHIN SPECIFICATION.

Description of Event or Problem · 1

A PERITONEAL DIALYSIS PT HAS REPORTED THAT DIALYSIS SOLUTION WAS LEAKING OUT OF THE CASSETTE AND INTO THE CYCLER. UPON REMOVING THE TUBING SET FOLLOWING TREATMENT, FLUID WAS DISCOVERED INSIDE THE CASSETTE COMPARTMENT OF THE CYCLER. PT HAD NO ILL EFFECTS. SAMPLE HAS NOT BEEN RETURNED TO THE MANUFACTURER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
31463 LIBERTY CYCLER SET, DUAL PATIENT CONENCT LIBERTY DIALYSIS CYCLER TUBING FKX REYNOSA MANUFACTURING 12JR08096

Patients

Seq Age Sex Outcome Treatment
1 LIBERTY DIALYSIS CYCLER