FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK SPIRIT

MDR report key: 2932136 · Received January 18, 2013

Report

Report Number
2183996-2013-00071
Event Type
Malfunction
Date Received
January 18, 2013
Date of Event
January 16, 2013
Report Date
January 16, 2013
Manufacturer
ROCHE HEALTH SOLUTIONS INC.
Product Code
LZG
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS INCIDENT OCCURRED OUTSIDE OF THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

THE COMPLAINT CANNOT BE VERIFIED DUE TO CARELESS HANDLING OF THE PRODUCT. THE SOFTCOMPONENTS OF THE HOUSING ARE WORN DOWN HEAVILY AND RESTRICTED HANDLING OF THE PUMP IS A POSSIBLE IMPLICATION. THEREFORE, MOISTURE ENTERED THE INSULIN PUMP AND DESTROYED THE FUNCTIONALITY OF THE BUTTONS AND THE PUMP ELECTRONICS. THE DOWN BUTTON IS PERMANENT ACTIVE DUE TO AN EXTERNAL MECHANIC DAMAGE. THE PROBLEM MENTIONED BY THE CUSTOMER IS RELATED TO CARELESS HANDLING OF THE PRODUCT.

Description of Event or Problem · 1

PATIENT REPORTED THE LATERAL BUTTON (DOWN BUTTON) ON THE INFUSION DEVICE IS STUCK AND EVEN THE OTHER BUTTONS DON'T WORK. PATIENT STATED THEY ARE ALL BLOCKED. PATIENT REPORTED THE RUBBER THAT COVERS THE LATERAL BUTTON DOWN CAME OFF. PATIENT STATED THE DISPLAY ON THE INFUSION DEVICE DOESN'T SHOW AN ALARM. PATIENT IS CURRENTLY USING THE BACKUP INFUSION DEVICE. PATIENT REPORTED SHE IS PREGNANT. NO FURTHER INFORMATION IS AVAILABLE. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. PRODUCT WAS REPLACED AND REQUESTED RETURN OF THE ALLEGED INFUSION DEVICE FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
28585 ACCU-CHEK SPIRIT LZG ROCHE HEALTH SOLUTIONS INC. 00700006862 NA

Patients

Seq Age Sex Outcome Treatment
1 RELATED ACCESSORIES| INSULIN INFUSION PUMP| INSULIN