FDA Adverse Event Malfunction Summary report: N

COULTER® ACT DIFF 12¿ ANALYZER

MDR report key: 2932039 · Received January 28, 2013

Report

Report Number
1061932-2013-00119
Event Type
Malfunction
Date Received
January 28, 2013
Date of Event
January 4, 2013
Report Date
January 4, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K973634
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) OBSERVED THE DILUENT FILTERS WERE CLOGGED AND GENERATED BACKPRESSURE WHICH WAS CAUSING THE PROBE TO LEAK. THE FSE REPLACED THE DILUENT FILTERS AND RESOLVED THE FLUID LEAK ISSUE. THE REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED APPROXIMATELY SEVERAL MILLILITERS OF BLOOD FLUID LEAKED FROM THE PROBE INVOLVING THE COULTER ACT DIFF 12 ANALYZER. THE CUSTOMER HAD ON PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES AND A LABORATORY COAT AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. PATIENT RESULTS WERE NOT IMPACTED. THE CUSTOMER HAS AN EXPOSURE CONTROL/RISK MANAGEMENT PLAN AT THE FACILITY. BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.