PRECISION®
Report
- Report Number
- 3006630150-2013-00157
- Event Type
- Malfunction
- Date Received
- January 28, 2013
- Date of Event
- September 10, 2012
- Report Date
- January 17, 2013
- Manufacturer
- BOSTON SCIENTIFIC NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- 030017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
Narratives
UPDATE TO INITIAL MDR FIELD: ADDITIONAL INFORMATION WAS RECEIVED THAT THERE WERE TWO LEADS DISPLAYING HIGH IMPEDANCES. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THIS TIME REGARDING THE LEADS.
THE COMPLAINT HAS BEEN CONFIRMED FOR S/N (B)(4). VISUAL INSPECTION REVEALED THAT THE LEAD WAS CLEANLY CUT APPROXIMATELY 11 INCHES FROM THE DISTAL END. X-RAY INSPECTION OF THE LEAD REVEALED THAT ALL OF THE CABLES WERE COMPLETELY BROKEN AT THE BENT/KINKED LOCATION OF THE LEAD. THE BENT/KINKED LOCATION IS 1 CM FROM THE SET SCREW MARK OF THE CLIK ANCHOR. IT APPEARS TO BE A RESULT OF FATIGUE. ELECTRICAL TESTS COULD NOT BE PERFORMED DUE TO BROKEN CABLES. THE COMPLAINT HAS BEEN CONFIRMED FOR S/N (B)(4). VISUAL INSPECTION REVEALED THAT THE LEAD WAS CLEANLY CUT APPROXIMATELY 14 INCHES FROM THE DISTAL END. X-RAY INSPECTION OF THE LEAD REVEALED THAT 2 CABLES WERE COMPLETELY BROKEN AT THE BENT/KINKED LOCATION OF THE LEAD. THE BENT/KINKED LOCATION IS 1 CM FROM THE SET SCREW MARK OF THE CLIK ANCHOR. IT APPEARS TO BE A RESULT OF FATIGUE. ELECTRICAL TESTS COULD NOT BE PERFORMED DUE TO BROKEN CABLES.
ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED: MODEL #: SC-2138-70 SERIAL/LOT#: (B)(4), DESCRIPTION: SCS 70CM III LEAD.
A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING INEFFECTIVE THERAPY. A SYSTEM CHECK REVEALED HIGH IMPEDANCES. THE PATIENT UNDERWENT A REVISION PROCEDURE AND THE PHYSICIAN CHOSE TO REPLACE THE LEAD. THE PATIENT IS REPORTEDLY DOING WELL FOLLOWING THE PROCEDURE.
A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING INEFFECTIVE THERAPY. A SYSTEM CHECK REVEALED HIGH IMPEDANCES. THE PATIENT UNDERWENT A REVISION PROCEDURE AND THE PHYSICIAN CHOSE TO REPLACE THE LEAD. THE PATIENT IS REPORTEDLY DOING WELL FOLLOWING THE PROCEDURE.
A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING INEFFECTIVE THERAPY. A SYSTEM CHECK REVEALED HIGH IMPEDANCES. THE PATIENT UNDERWENT A REVISION PROCEDURE AND THE PHYSICIAN CHOSE TO REPLACE THE LEAD. THE PATIENT IS REPORTEDLY DOING WELL FOLLOWING THE PROCEDURE.
A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING INEFFECTIVE THERAPY. A SYSTEM CHECK REVEALED HIGH IMPEDANCES. THE PATIENT UNDERWENT A REVISION PROCEDURE AND THE PHYSICIAN CHOSE TO REPLACE THE LEAD. THE PATIENT IS REPORTEDLY DOING WELL FOLLOWING THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 37434 | PRECISION® | SPINAL CORD STIMULATOR | LGW | BOSTON SCIENTIFIC NEUROMODULATION | SC-2138-70 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |