FDA Adverse Event Malfunction Summary report: N

GCX

MDR report key: 2931798 · Received October 17, 2012

Report

Report Number
2931798
Event Type
Malfunction
Date Received
October 17, 2012
Date of Event
October 17, 2012
Report Date
December 17, 2012
Manufacturer
GCX CORPORATION
Product Code
MHX
Product Problem
Yes
Report Source
User Facility report
Reporter Location
SC, US

Narratives

Description of Event or Problem · 1

THE GCX MOUNTING BRACKET ON THE SPACELAB HOUSING BROKE AND THE HOUSING AND THE MODULES FELL. THE UNIT IS VERY HEAVY AND IS LOCATED ABOVE THE PATIENT'S HEAD. THE NURSE WAS UNSURE OF WHAT CAUSED THE MOUNT TO BREAK AS NO ONE WAS MOVING OR TOUCHING IT AT THE TIME IT GAVE WAY. THIS IS THE 4TH MODULE BRACKET TO BREAK IN THE CRITICAL CARE UNIT. THE MOUNTING ARMS THAT SUPPORTS THE SPACELAB MODULE HOUSING BROKE JUST ABOVE THE HOUSING AT THE SWIVEL POINT.THE SERVICE REP FOR GCX CORPORATION CAME TO THE FACILITY AND REINFORCED THE CONNECTION BETWEEN THE ARM AND THE MONITOR WITH PLASTIC TIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GCX MONITOR, PHYSIOLOGICAL, MOUNTING ARM MHX GCX CORPORATION * *

Patients

Seq Age Sex Outcome Treatment
1 *