HAKIM PROGRAMMABLE VALVE WITH PRECHAMBER
Report
- Report Number
- 1226348-2013-11582
- Event Type
- Injury
- Date Received
- January 28, 2013
- Date of Event
- December 19, 2012
- Manufacturer
- CODMAN & SHURTLEFF, INC
- Product Code
- JXG
- PMA / PMN Number
- PK974739
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
UPON COMPLETION OF THE INVESTIGATION, A FOLLOW UP REPORT WILL BE FILED.
EXAMINATION OF THE VALVE REVEALED THE PRESENCE OF BIOLOGICAL DEBRIS WITHIN THE DEVICE. THIS BIOLOGICAL DEBRIS CAUSED AN OCCLUSION IN THE DEVICE AND FOR THE RETURNED VALVE TO FAIL THE PROGRAMMING TEST, WHICH APPEARS TO HAVE CAUSED THE DIFFICULTY ENCOUNTERED BY THE CUSTOMER. REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE VALVE MET SPECIFICATION REQUIREMENTS WHEN RELEASED TO STOCK. BASED ON THE RESULTS OF THIS INVESTIGATION NO FURTHER ACTION IS REQUIRED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME THIS COMPLAINT IS CLOSED.
IN 2008 THE PATIENT DEVELOPED MENINGITIS. AS A RESULT THE PATIENT RECEIVED A NEW VALVE, HOWEVER IT IS NOW BEING REPORTED THAT THE VALVE IS NOT LONGER ADJUSTABLE. AS A RESULT THE DEVICE WAS REVISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 36406 | HAKIM PROGRAMMABLE VALVE WITH PRECHAMBER | SHUNT, CENTRAL NERVOUS SYSTEM & COMPS | JXG | CODMAN & SHURTLEFF, INC | CGMB3J |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Required Intervention |