FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2931717 · Received January 28, 2013

Report

Report Number
1416980-2013-02076
Event Type
Malfunction
Date Received
January 28, 2013
Date of Event
January 1, 2013
Report Date
January 3, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF "AIR ALARM" WAS CONFIRMED DURING DEVICE EVALUATION AS ALARM F-1. THE ROOT CAUSE OF THE PROBLEM WAS AIR LINE SENSOR - OUT OF CALIBRATION. THE AIR IN LINE SENSOR WAS RECALIBRATED TO CORRECT THE REPORTED CONDITION.

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON-SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER (B)(4) THAT A FLOGARD INFUSION PUMP HAD AN "AIR ALARM." IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT; THEREFORE, NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
37710 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1