FLOGARD
Report
- Report Number
- 1416980-2013-02076
- Event Type
- Malfunction
- Date Received
- January 28, 2013
- Date of Event
- January 1, 2013
- Report Date
- January 3, 2013
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF "AIR ALARM" WAS CONFIRMED DURING DEVICE EVALUATION AS ALARM F-1. THE ROOT CAUSE OF THE PROBLEM WAS AIR LINE SENSOR - OUT OF CALIBRATION. THE AIR IN LINE SENSOR WAS RECALIBRATED TO CORRECT THE REPORTED CONDITION.
(B)(4).
(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON-SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
IT WAS REPORTED TO BAXTER (B)(4) THAT A FLOGARD INFUSION PUMP HAD AN "AIR ALARM." IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT; THEREFORE, NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.