ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-01244
- Event Type
- Malfunction
- Date Received
- January 28, 2013
- Report Date
- January 22, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP #1 DATE OF SUBMISSION 04/03/2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS ARE APPROPRIATELY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL KEY CONTACTS. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED A CRACKED BATTERY COMPARTMENT, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION. THE USER GUIDE WARNS THAT CRACKS, CHIPS, OR DAMAGE TO THE PUMP MAY IMPACT THE BATTERY CONTACT AND/OR THE WATERPROOF FEATURE OF THE PUMP. INVESTIGATION ALSO REVEALED A SCRATCHED DISPLAY LENS FILM. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE PATIENT CONTACTED ANIMAS ON (B)(6) 2013, REPORTING THAT ALL KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE PATIENT CONFIRMED THAT THE KEYPAD WAS INTACT AND NO EVIDENCE OF MOISTURE OR CORROSION NOTED. THE PATIENT STATED THAT THERE WAS NO MOISTURE TO THE PUMP. THE PATIENT REPORTEDLY WORE THE PUMP ATTACHED TO A BELT AND DOES NOT CLEAN THE PUMP. THERE IS NO REPORTED ADVERSE EVENT WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON THE ALLEGATION OF THE KEYPAD RESPONSE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 37544 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |