FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2931594 · Received January 28, 2013

Report

Report Number
2531779-2013-01244
Event Type
Malfunction
Date Received
January 28, 2013
Report Date
January 22, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 DATE OF SUBMISSION 04/03/2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS ARE APPROPRIATELY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL KEY CONTACTS. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED A CRACKED BATTERY COMPARTMENT, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION. THE USER GUIDE WARNS THAT CRACKS, CHIPS, OR DAMAGE TO THE PUMP MAY IMPACT THE BATTERY CONTACT AND/OR THE WATERPROOF FEATURE OF THE PUMP. INVESTIGATION ALSO REVEALED A SCRATCHED DISPLAY LENS FILM. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE PATIENT CONTACTED ANIMAS ON (B)(6) 2013, REPORTING THAT ALL KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE PATIENT CONFIRMED THAT THE KEYPAD WAS INTACT AND NO EVIDENCE OF MOISTURE OR CORROSION NOTED. THE PATIENT STATED THAT THERE WAS NO MOISTURE TO THE PUMP. THE PATIENT REPORTEDLY WORE THE PUMP ATTACHED TO A BELT AND DOES NOT CLEAN THE PUMP. THERE IS NO REPORTED ADVERSE EVENT WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON THE ALLEGATION OF THE KEYPAD RESPONSE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
37544 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 11 YR