FDA Adverse Event Malfunction Summary report: N

TRIATHLON #4 CR INSERT 9MM

MDR report key: 2931345 · Received January 28, 2013

Report

Report Number
0002249697-2013-00358
Event Type
Malfunction
Date Received
January 28, 2013
Date of Event
January 14, 2013
Report Date
January 14, 2013
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
JWH
PMA / PMN Number
K040267
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THERE ARE SIGNIFICANT MARKS AND INDENTATIONS ON THE INSERT COMPONENT INCLUDING MARKS ON THE INFERIOR AND POSTERIOR SURFACES OF THE COMPONENT, WHICH IS EVIDENCE THAT THE INSERT COMPONENT CAME INTO CONTACT WITH SOME FORM OF RESISTANCE DURING THE ATTEMPTED POSITIONING AND LOCKING OF THE INSERT INTO THE BASEPLATE COMPONENT. THE EVENT AS REPORTED WAS CONFIRMED. DHR REVIEW FOR THE REPORTED LOT WAS SATISFACTORY. CHR REVIEW FOR THE REPORTED LOT CONFIRMED THAT THERE WERE NO SIMILAR EVENTS REPORTED FOR THE LOT. THE INVESTIGATION CONCLUDED THAT THE SEATING ISSUE WAS CAUSED BY AN OBSTRUCTION ENCOUNTERED BY THE INSERT COMPONENT WHILE TRYING TO ENGAGE THE COMPONENT WITH THE BASEPLATE.

Additional Manufacturer Narrative · 1

WHEN COMPLETED, THE EVALUATION SUMMARY WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

WHEN POLY WAS INSERTED IT WOULDN'T SIT DOWN ON THE MEDIAL ANTERIOR SIDE. THERE WAS NOTHING IN ITS WAY (IE NO CEMENT CAUGHT IN TRAY). IT WAS IF THE WIRE WOULDN¿T GO PAST THE LITTLE TAB ON THE ANTERIOR LIP OF THE TRAY. IT WAS STILL WOULDN'T UP AND DOWN ANTEROMEDIALLY, IT WAS REMOVED AND A NEW POLY WAS PUT IN, WHICH LOCKED DOWN WITH NO DRAMAS.

Description of Event or Problem · 1

WHEN POLY WAS INSERTED IT WOULDN'T SIT DOWN ON THE MEDIAL ANTERIOR SIDE. THERE WAS NOTHING IN ITS WAY (IE, NO CEMENT CAUGHT IN TRAY). IT WAS AS IF THE WIRE WOULDN'T GO PAST THE LITTLE TAB ON THE ANTERIOR LIP OF THE TRAY. IT WAS STILL POSITIONING UP AND DOWN ANTEROMEDIALLY, IT WAS REMOVED AND A NEW POLY WAS PUT IN, WHICH LOCKED DOWN WITH NO DRAMAS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
36961 TRIATHLON #4 CR INSERT 9MM IMPLANT JWH STRYKER ORTHOPAEDICS-MAHWAH LDA383

Patients

Seq Age Sex Outcome Treatment
1 65 YR