ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-01220
- Event Type
- Malfunction
- Date Received
- January 28, 2013
- Report Date
- January 19, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP #1 DATE OF SUBMISSION 03/21/2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: BLACK BOX REVIEW SHOWS NUMEROUS UNUSUAL POWER ON RESET. THE BATTERY CAP AND BATTERY COMPARTMENT THREADS WERE INTACT AND NO CRACKS OBSERVED IN THE BATTERY COMPARTMENT. THE BATTERY CAP HEIGHT INCH WAS ABOVE SPECIFICATIONS AND THE WIDTH INCH WITHIN SPECIFICATIONS. THE BATTERY CAP IS ABLE TO MAINTAIN AN ELECTRICAL CONNECTION, THE PUMP POWERS ON NORMAL WITH NO ALARMS. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED A DIM DISPLAY SCREEN.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE REPORTER CONTACTED ANIMAS ON (B)(6) 2013 REPORTING THAT THE PUMP REBOOTED TWO TIMES IN 15 MINUTES. HE STATED THAT EACH TIME THEY WILL HEAR THE PUMP BEEP AND VERIFY SCREEN CAME UP. THE REWIND, LOAD AND PRIMES STEPS WERE COMPLETED. THE REPORTER CONFIRMED THAT THERE WAS NO DAMAGE TO THE BATTERY COMPARTMENT OR BATTERY CAP AND THE BATTERY CAP SECURES PROPERLY TO PUMP, NO YELLOW SHOWING. THERE IS NO REPORTED ADVERSE EVENT WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON THE ALLEGATION OF THE INTERMITTENT POWER ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 36831 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |