FDA Adverse Event Malfunction Summary report: N

5MM, 33CM PEEK MONOPOLAR HANDLE 33CM

MDR report key: 2931206 · Received January 28, 2013

Report

Report Number
0002936485-2013-00002
Event Type
Malfunction
Date Received
January 28, 2013
Date of Event
January 4, 2013
Report Date
January 4, 2013
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
KOG
PMA / PMN Number
K973259
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RETURNED FOR INVESTIGATION. THE REPORTED FAILURE MODE WAS CONFIRMED. THE UNIT FAILED THE INSULATION SCAN TEST. UNDER THE MICROSCOPE, THERE IS A SMALL SCRATCH ON THE INSULATION LOCATED AT THE PROXIMAL END. POSSIBLE ROOT CAUSE(S) FOR THE COMPROMISED INSULATION ARE: MULTIPLE USES OF FLASH STERILIZATION, RAPID COOLING AFTER STERILIZATION AND/OR CONTACT WITH METAL TRAY DURING STERILIZATION. AN ACTION HAS BEEN OPENED TO ADDRESS THE INSULATION FAILURE FOR OUR OEM SUPPLIER FOR OUR LAP INSTRUMENTS USING POLYMER RESIN INSULATION. IN SUM, THE PRODUCT WAS RETURNED FOR INVESTIGATION AND THE REPORTED FAILURE WAS CONFIRMED. THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION ON THE UNIT HAS BEEN COMPROMISED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION ON THE UNIT HAS BEEN COMPROMISED.