5MM, 33CM PEEK MONOPOLAR HANDLE 33CM
Report
- Report Number
- 0002936485-2013-00002
- Event Type
- Malfunction
- Date Received
- January 28, 2013
- Date of Event
- January 4, 2013
- Report Date
- January 4, 2013
- Manufacturer
- STRYKER ENDOSCOPY-SAN JOSE
- Product Code
- KOG
- PMA / PMN Number
- K973259
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.
THE PRODUCT WAS RETURNED FOR INVESTIGATION. THE REPORTED FAILURE MODE WAS CONFIRMED. THE UNIT FAILED THE INSULATION SCAN TEST. UNDER THE MICROSCOPE, THERE IS A SMALL SCRATCH ON THE INSULATION LOCATED AT THE PROXIMAL END. POSSIBLE ROOT CAUSE(S) FOR THE COMPROMISED INSULATION ARE: MULTIPLE USES OF FLASH STERILIZATION, RAPID COOLING AFTER STERILIZATION AND/OR CONTACT WITH METAL TRAY DURING STERILIZATION. AN ACTION HAS BEEN OPENED TO ADDRESS THE INSULATION FAILURE FOR OUR OEM SUPPLIER FOR OUR LAP INSTRUMENTS USING POLYMER RESIN INSULATION. IN SUM, THE PRODUCT WAS RETURNED FOR INVESTIGATION AND THE REPORTED FAILURE WAS CONFIRMED. THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.
IT WAS REPORTED THAT THE INSULATION ON THE UNIT HAS BEEN COMPROMISED.
IT WAS REPORTED THAT THE INSULATION ON THE UNIT HAS BEEN COMPROMISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 37664 | 5MM, 33CM PEEK MONOPOLAR HANDLE 33CM | ENDOSCOPE AND/OR ACCESSORIES | KOG | STRYKER ENDOSCOPY-SAN JOSE | 1144707D |