FDA Adverse Event Malfunction Summary report: N

PRECISION®

MDR report key: 2930759 · Received January 28, 2013

Report

Report Number
3006630150-2013-00154
Event Type
Malfunction
Date Received
January 28, 2013
Date of Event
January 10, 2013
Report Date
January 10, 2013
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION
Product Code
LGW
PMA / PMN Number
030017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION INDICATED THAT THE IPG PASSED VISUAL, ELECTRICAL, PERFORMANCE AND PHOTOGRAPHIC IMAGING TESTS PERFORMED. THE COMPLAINT OF DIFFICULTY CHARGING WAS NOT CONFIRMED. CHARGE PROFILE REVEALED VARIOUS TIMES OF ERRATIC COUPLING/POOR ALIGNMENT OF THE CHARGER TO THE IPG. THE REASON FOR THE ERRATIC COUPLING IS UNKNOWN. BATTERY PROFILE REVEALED A MAXIMUM DEPLETION RATE OF 11MV PER DAY, WHICH IS WITHIN THE EXPECTED RANGE. DEVICE EXHIBITS NORMAL CHARGING AND DISCHARGING CHARACTERISTICS.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT THE PATIENT UNDERWENT AN IPG REPLACEMENT PROCEDURE DUE TO INABILITY TO HOLD ITS CHARGE. THE PATIENT RECEIVED GOOD STIMULATION COVERAGE POSTOPERATIVELY. THE PATIENT HAD A PREVIOUS NONDEVICE RELATED SURGERY WHEREIN ELECTROCAUTERY WAS USED. MONOPOLAR ELECTROCAUTERY IS A KNOWN SOURCE OF HIGH VOLTAGE TRANSIENT SIGNAL AND CURRENT COMPANY LABELING WARNS AGAINST THE USE OF MONOPOLAR ELECTROCAUTERY (PHYSICIAN'S IMPLANT MANUAL 9055940-001).

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT THE PATIENT UNDERWENT AN IPG REPLACEMENT PROCEDURE DUE TO INABILITY TO HOLD ITS CHARGE. THE PATIENT RECEIVED GOOD STIMULATION COVERAGE POSTOPERATIVELY. THE PATIENT HAD A PREVIOUS NON-DEVICE RELATED SURGERY WHEREIN ELECTROCAUTERY WAS USED. MONOPOLAR ELECTROCAUTERY IS A KNOWN SOURCE OF HIGH VOLTAGE TRANSIENT SIGNAL AND CURRENT COMPANY LABELING WARNS AGAINST THE USE OF MONOPOLAR ELECTROCAUTERY (PHYSICIAN'S IMPLANT MANUAL 9055940-001).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
37524 PRECISION® SPINAL CORD STIMULATOR LGW BOSTON SCIENTIFIC NEUROMODULATION SC-1110-02 NA

Patients

Seq Age Sex Outcome Treatment
1 43 YR