FDA Adverse Event Malfunction Summary report: N

RESERVOIR 3ML

MDR report key: 2930689 · Received January 28, 2013

Report

Report Number
3004209178-2013-90591
Event Type
Malfunction
Date Received
January 28, 2013
Date of Event
January 15, 2013
Report Date
January 15, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K032005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSPECTED ONE OPENED AND USED RESERVOIR. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIR PASSED PER SPECIFICATION. NO LEAKAGE ANOMALY WAS OBSERVED DURING ANALYSIS. CHECKED O-RINGS FOR DEFECTS AND NONE WERE FOUND.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT INSULIN FROM BACK OF THE RESERVOIR LEAKED PAST THE O-RINGS. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
36429 RESERVOIR 3ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-332A

Patients

Seq Age Sex Outcome Treatment
1 41 YR