PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 2930624
·
Received January 28, 2013
Report
- Report Number
- 3004209178-2013-90604
- Event Type
- Malfunction
- Date Received
- January 28, 2013
- Date of Event
- January 2, 2013
- Report Date
- January 22, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WY
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP PASSED THE SELF TEST. HOWEVER, THE DEVICE FAILED THE DISPLACEMENT TEST AS A RESULT OF THE MOTOR ENCODER SIGNAL BEING OUT OF PHASE. FURTHER TESTING COULD NOT BE PERFORMED DUE TO THE ANOMALY.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP HAD ALARMED MULTIPLE TIMES DURING NORMAL DEVICE USE. THE BLOOD GLUCOSE READING WAS 346MG/DL. TROUBLESHOOTING WAS PERFORMED. REVIEWED THE ALARM HISTORY AND FOUND SEVERAL DIFFERENT ERROR ALARMS. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED. NO FURTHER INFORMATION WAS PROVIDED.