FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 2930624 · Received January 28, 2013

Report

Report Number
3004209178-2013-90604
Event Type
Malfunction
Date Received
January 28, 2013
Date of Event
January 2, 2013
Report Date
January 22, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP PASSED THE SELF TEST. HOWEVER, THE DEVICE FAILED THE DISPLACEMENT TEST AS A RESULT OF THE MOTOR ENCODER SIGNAL BEING OUT OF PHASE. FURTHER TESTING COULD NOT BE PERFORMED DUE TO THE ANOMALY.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP HAD ALARMED MULTIPLE TIMES DURING NORMAL DEVICE USE. THE BLOOD GLUCOSE READING WAS 346MG/DL. TROUBLESHOOTING WAS PERFORMED. REVIEWED THE ALARM HISTORY AND FOUND SEVERAL DIFFERENT ERROR ALARMS. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
37998 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAH

Patients

Seq Age Sex Outcome Treatment
1 21 YR