FDA Adverse Event Malfunction Summary report: N

CONSTRAINED LINER INSERTER IMPACTOR TIP 22MM

MDR report key: 2930488 · Received January 28, 2013

Report

Report Number
0002249697-2013-00334
Event Type
Malfunction
Date Received
January 28, 2013
Date of Event
January 8, 2013
Report Date
January 8, 2013
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

WHEN COMPLETED, THE EVALUATION SUMMARY WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

DHR INDICATES THAT THE REPORTED DEVICES WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. CHR INDICATED THAT THERE WERE NO SIMILAR EVENTS FOR THE REPORTED LOT. VISUAL INSPECTION OF RETURNED DEVICE CONFIRMED THE EVENT. A LARGE SHARD OF PLASTIC FRACTURED FROM DEVICE IN REGION OF THREADED MATING HOLE. DAMAGE TO DEVICE SUGGESTS MATING IMPACTOR HANDLE WAS NOT FULLY ATTACHED TO SUBJECT DEVICE PRIOR TO IMPACTION. MATERIAL ANALYSIS FOUND NO MATERIAL OR MANUFACTURING DEFECTS ON FRACTURE SURFACES EXAMINED. THE INVESTIGATION CONCLUDED THAT IMPACTOR TIP DAMAGE WAS CAUSED BY USER MISUSE. THE IMPACTOR TIP WAS NOT FULLY THREADED ON IMPACTOR HANDLE BEFORE USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT SURGEON WAS DOING A REVISION FOR A DISLOCATING/DISASSOCIATING RIGHT HIP. SURGEON REMOVED LINER WHICH WAS DISASSOCIATING FROM CUP AND WHILE SURGEON WAS IMPLANTING A NEW CONSTRAINED LINER, WHILE USING THE LINER IMPACTION TOOL, THE TOOL BROKE - A LARGE PORTION OF THE IMPACTOR BROKE OFF AND WOULD NO LONGER THREAD INTO THE HANDLE. SURGEON USED THE LARGER HEAD IMPACTOR AND COMPLETED THE CASE WITHOUT ANY DELAY OR INTERRUPTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT SURGEON WAS DOING A REVISION FOR A DISLOCATING/DISASSOCIATING RIGHT HIP. SURGEON REMOVED LINER WHICH WAS DISASSOCIATING FROM CUP AND WHILE SURGEON WAS IMPLANTING A NEW CONSTRAINED LINER, WHILE USING THE LINER IMPACTION TOOL, THE TOOL BROKE - A LARGE PORTION OF THE IMPACTOR BROKE OFF AND WOULD NO LONGER THREAD INTO THE HANDLE. SURGEON USED THE LARGER HEAD IMPACTOR AND COMPLETED THE CASE WITHOUT ANY DELAY OR INTERRUPTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
37970 CONSTRAINED LINER INSERTER IMPACTOR TIP 22MM INSTRUMENT LXH STRYKER ORTHOPAEDICS-MAHWAH KPWMED

Patients

Seq Age Sex Outcome Treatment
1 71 YR Other