FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2930486 · Received January 28, 2013

Report

Report Number
1416980-2013-02027
Event Type
Malfunction
Date Received
January 28, 2013
Date of Event
January 1, 2013
Report Date
January 3, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON-SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF "F-94 ALARM" WAS CONFIRMED DURING DEVICE EVALUATION. THE ROOT CAUSE WAS DETERMINED TO BE A DAMAGED MAIN BATTERY. THE MAIN BATTERY WAS REPLACED TO RESOLVE THE REPORTED CONDITION. A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS CUSTOMER SERVICE REQUEST ON THIS PUMP.

Additional Manufacturer Narrative · 1

(B)(4)

Description of Event or Problem · 1

BAXTER (B)(6) REPORTED A FLOGARD INFUSION PUMP WITH A F-94 ALARM. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT; THEREFORE, NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.