SYSMEX CA-1500 ANALYZER
Report
- Report Number
- 1422681-2013-00001
- Event Type
- Injury
- Date Received
- January 22, 2013
- Report Date
- December 24, 2012
- Manufacturer
- SYSMEX CORPORATION
- Product Code
- GKP
- PMA / PMN Number
- K993299
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE SYSMEX CA-1500 ANALYZER PERFORMED AS EXPECTED; NO MALFUNCTION WAS REPORTED. THE OPERATOR REACHED UNDER THE LIGHT SHIELD WHILE THE ANALYZER WAS IN OPERATION IN AN ATTEMPT TO ADD REAGENT. A REVIEW OF THE SYSMEX CA-1500 SERIES INSTRUCTIONS FOR USE INCLUDES SEVERAL USER WARNINGS TO NOT OPEN THE LIGHT SHIELD COVER DURING ANALYSIS. THIS COULD CAUSE INJURY. NONE OF THE COMPLAINT NARRATIVE STATES OR IMPLIES THAT THE DEVICE FAILED IN ANY WAY. CURRENT STATUS OF THE OPERATOR IS NOT KNOWN. SPECIFIC CORRECTIVE ACTION ON THE INSTRUMENT IS NOT KNOWN.
AN INTERNAL INCIDENT REPORT WAS PREPARED AT A LABORATORY IN THE UNITED STATES AND SUBMITTED TO SIEMENS HEALTHCARE DIAGNOSTICS ON (B)(4) 2012 REGARDING THE OPERATOR OF A SYSMEX CA-1500 ANALYZER, SERIAL NUMBER (B)(4), WHO REACHED UNDER THE LIGHT SHIELD AND ATTEMPTED TO ADD REAGENT WHILE THE ANALYZER WAS IN OPERATION. HER HAND WAS STRUCK AND CUT BY THE MOVING REAGENT PROBE. THE PROBE BODY WAS CRACKED AND BROKEN BECAUSE OF THE IMPACT. IT IS NOT KNOWN IF SHE WAS WEARING APPROPRIATE PERSONNEL PROTECTIVE EQUIPMENT (PPE), SUCH AS GLOVES, AT THE TIME. THE OPERATOR WAS TREATED AT THE EMERGENCY ROOM AND WAS GIVEN ANTI-VIRAL PROPHYLAXIS. THE HAND WAS CLEANED. SHE WAS DRAWN FOR BASELINE CBC AND CHEMISTRY PROFILE. IT IS NOT KNOWN IF THE OPERATOR OR POTENTIAL SOURCE WAS TESTED FOR EXPOSURE TO BLOOD-BORNE PATHOGENS. THE OCCURRENCE WAS REPORTED TO SYSMEX CORPORATION (B)(4) ON (B)(4), 2012 AND RELAYED TO (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 31172 | SYSMEX CA-1500 ANALYZER | AUTOMATED BLOOD COAGULATION ANALYZER | GKP | SYSMEX CORPORATION | CA-1500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Required Intervention |