FDA Adverse Event Injury Summary report: N

REJUVENATE STRGHT PRFIT TMZF MOD STEM SIZE 7

MDR report key: 2930247 · Received January 25, 2013

Report

Report Number
0002249697-2013-00304
Event Type
Injury
Date Received
January 25, 2013
Date of Event
June 8, 2012
Report Date
June 8, 2012
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
MEH
PMA / PMN Number
K071082
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AN EVALUATION OF THE DEVICE CANNOT BE PERFORMED AS THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER. ADDITIONAL INFORMATION PERTAINING TO THE DEVICE REFERENCED IN THIS REPORT (INCLUDING X-RAYS AND MEDICAL RECORDS) HAS BEEN REQUESTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THE EVALUATION SUMMARY WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

AN EVENT REGARDING ALTR INVOLVING A REJUVENATE MODULAR DEVICE WAS REPORTED. THE EVENT WAS CONFIRMED. DEVICE EVALUATION WAS NOT PERFORMED AS NO DEVICES WERE RECEIVED. A DEVICE HISTORY REVIEW FOUND THAT THE DEVICES IN THE REPORTED LOT WERE ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. THE SUPER AND CHS COMPLAINT DATABASES SHOW THERE HAVE BEEN OTHER EVENTS FOR THE REPORTED LOT. SIMILAR EVENTS HAVE OCCURRED FOR THE CATALOG NUMBER AND PRODUCT FAMILY. THESE EVENTS WERE DETERMINED TO BE ASSOCIATED WITH RA 2012-067. VOLUNTARY RECALL RA 2012-067 WAS INITIATED FOR ABGII AND REJUVENATE MODULAR STEMS AND NECKS DUE TO THE POTENTIAL RISKS ASSOCIATED WITH THESE DEVICES. THE REPORTED ALTR IS CONSIDERED TO BE UNDER THE SCOPE OF THIS RECALL. NO FURTHER INVESTIGATION IS REQUIRED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED WITH INCREASED PAIN IN JOINT AREA, ELEVATED CHROME AND COBALT SERUM LEVELS, POSITIVE MARS SCAN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED WITH INCREASED PAIN IN JOINT AREA, ELEVATED CHROME AND COBALT SERUM LEVELS, POSITIVE MARS SCAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
35244 REJUVENATE STRGHT PRFIT TMZF MOD STEM SIZE 7 IMPLANT MEH STRYKER ORTHOPAEDICS-MAHWAH MJTWM9

Patients

Seq Age Sex Outcome Treatment
1 Other