MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT
Report
- Report Number
- 2955842-2013-00286
- Event Type
- Malfunction
- Date Received
- January 25, 2013
- Date of Event
- December 24, 2012
- Report Date
- December 26, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. THE CUSTOMER INITIALLY REPORTED A FRAYED WIRE, HOWEVER, FOR CLARIFICATION THE DAMAGED INSTRUMENT COMPONENT WAS A BROKEN GRIP CABLE. BASED ON THIS CLARIFICATION, THE CUSTOMER REPORTED COMPLAINT IS CONFIRMED. ONE GRIP CLOSE CABLE IS BROKEN AT THE DISTAL IDLERS. IDLER PULLEY SPINS FREELY AND WAS NOT DAMAGED. CABLE SEGMENT STICKS OUT AT WRIST. OTHER CABLES AT WRIST ARE NOT DAMAGED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS OBSERVED DURING CENTRAL PROCESSING THAT THE MEGA SUTURE CUT NEEDLE DRIVER INSTRUMENT HAD A FRAYED WIRE. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.