FDA Adverse Event Injury Summary report: N

LRG TAP PRI MOD NCK 0DEG 34MM

MDR report key: 2930209 · Received January 25, 2013

Report

Report Number
0002249697-2013-00285
Event Type
Injury
Date Received
January 25, 2013
Date of Event
October 23, 2012
Report Date
January 9, 2013
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
MEH
PMA / PMN Number
K092561
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AN EVALUATION OF THE DEVICE CANNOT BE PERFORMED AS THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, IT WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

UPDATED STERILE LOT NUMBER BASED ON ADDITIONAL INFORMATION. THE PATIENT IS (B)(6). AN EVENT REGARDING EXCESSIVE METAL IONS INVOLVING A REJUVENATE MODULAR DEVICE WAS REPORTED. THE EVENT WAS CONFIRMED. REVIEW OF DEVICE HISTORY RECORDS FOUND THE DEVICES IN THE REPORTED LOT WERE ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. THE COMPLAINT DATABASES SHOW THERE HAVE BEEN OTHER EVENTS FOR THE REPORTED LOT. SIMILAR EVENTS HAVE OCCURRED FOR THE CATALOG NUMBER AND PRODUCT FAMILY. THESE EVENTS WERE DETERMINED TO BE ASSOCIATED WITH RA 2012-067. VOLUNTARY RECALL RA 2012-067 WAS INITIATED FOR ABGII AND REJUVENATE MODULAR STEMS AND NECKS DUE TO THE POTENTIAL RISKS ASSOCIATED WITH THESE DEVICES. THE REPORTED EXCESSIVE METAL IONS IS CONSIDERED TO BE UNDER THE SCOPE OF THIS RECALL.

Description of Event or Problem · 1

LABS TAKEN ON (B)(6) 2012 INDICATE ELEVATED COBALT (11 NG/ML) AND CHROMIUM (1.2 NG/ML) ION LEVELS; NO FURTHER INFORMATION AVAILABLE IN CLINIC OR HOSPITAL RECORDS. THE PATIENT IS DUE FOR A 2-YEAR FOLLOW-UP VISIT IN (B)(6) 2013.

Description of Event or Problem · 1

LABS TAKEN ON (B)(6) 2012 INDICATE ELEVATED COBALT (11 NG/ML) AND CHROMIUM (1.2 NG/ML) ION LEVELS; NO FURTHER INFORMATION AVAILABLE IN CLINIC OR HOSPITAL RECORDS. THE PATIENT IS DUE FOR A 2-YEAR FOLLOW-UP VISIT IN (B)(6) 2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
35207 LRG TAP PRI MOD NCK 0DEG 34MM IMPLANT MEH STRYKER ORTHOPAEDICS-MAHWAH 34887002

Patients

Seq Age Sex Outcome Treatment
1 73 YR Other