FDA Adverse Event Malfunction Summary report: N

MEGA SUTURECUT NEEDLE DRIVER

MDR report key: 2930199 · Received January 25, 2013

Report

Report Number
2955842-2013-00294
Event Type
Malfunction
Date Received
January 25, 2013
Date of Event
November 28, 2012
Report Date
December 28, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. THE CUSTOMER INITIALLY REPORTED A LOOSE WIRE ISSUE. HOWEVER, FOR CLARIFICATION THE DAMAGED INSTRUMENT COMPONENT WAS A BROKEN GRIP CABLE. BASED ON THIS CLARIFICATION, THE CUSTOMER REPORTED COMPLAINT IS CONFIRMED. ONE GRIP CLOSE CABLE WAS FOUND TO BE BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY SPUN FREELY AND WAS NOT DAMAGED. A CABLE SEGMENT WAS FOUND TO BE STICKING OUT AT THE WRIST. OTHER CABLES AT THE WRIST WERE NOT DAMAGED. ADDITIONAL FINDING: THE INSTRUMENT WAS FOUND TO HAVE BROKEN HOUSING SNAPS. FOUR OF FOUR HOUSING SNAP TABS AND THREE ADJACENT HOUSING PINS WERE BROKEN. THE HOUSING WAS ABLE TO BE REMOVED AS A RESULT OF THE DAMAGE. INSTRUMENT MAY HAVE BEEN MISHANDLED. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT A MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT HAD LOOSE WIRING. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
35600 MEGA SUTURECUT NEEDLE DRIVER ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420209-06 S10081020 289

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES