PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2013-00406
- Event Type
- Injury
- Date Received
- January 25, 2013
- Date of Event
- November 10, 2012
- Report Date
- January 4, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). THE OTHER PROGLIDE DEVICE REFERENCED IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER. EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. ANALYSIS OF THE RETURNED DEVICE SUBSTANTIATED THE REPORTED EXPERIENCE AS IT INDICATED THAT THE POSTERIOR CUFF MISS AND SUBSEQUENT ANTERIOR CUFF-TO-NEEDLE TIP DETACHMENT OCCURRED. CUFF TABS MEASURE ONE ONE-HUNDREDTH (0.01) OF AN INCH SQUARE. DUE TO THE EXTREMELY SMALL SIZE, A MISSING CUFF TAB IS NOT VISIBLE WITHOUT MAGNIFICATION AND WOULD NOT BE NOTED BY THE USER. THE PROBABLE CAUSE FOR THE REPORTED EVENT WAS DETERMINED TO BE RELATED TO THE OPERATIONAL CONTEXT DURING USE. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, A CUFF MISS OCCURRED. A SECOND PROGLIDE DEVICE WAS USED WITH THE SAME RESULTS. MANUAL ARTERIAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THE PHYSICIAN IS REPORTED TO BE TRAINED IN THE USE OF THE PROGLIDE DEVICE. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE ENTIRE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 35160 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 20810J1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | SHEATH: 6FR |