FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 2930085 · Received January 25, 2013

Report

Report Number
1416980-2013-01995
Event Type
Malfunction
Date Received
January 25, 2013
Date of Event
April 2, 2011
Report Date
January 15, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
Removal / Correction Number
1423500-01/08/10-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS PRODUCT IS NOT DISTRIBUTED IN THE US AND DOES NOT HAVE A 510(K) NUMBER. EVALUATION: THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS USE ERROR, TIDAL TOTAL UF REMOVAL SET TOO LOW. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2011, 07:29:56. DURING NIGHT DRAIN CYCLE 11, THE PATIENT'S ULTRAFILTRATION (UF) READING WAS 1093ML, INDICATING THE HOME PATIENT (HP) DRAINED 1093ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1000ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
35678 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1