FDA Adverse Event Malfunction Summary report: N

MAXSTAR UNIT MOUNT LIGHT

MDR report key: 2926477 · Received January 16, 2013

Report

Report Number
1017522-2013-00002
Event Type
Malfunction
Date Received
January 16, 2013
Date of Event
December 20, 2012
Report Date
December 21, 2012
Manufacturer
MARUS
Product Code
EAZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE LENS HEAT SHIELD/HOLDER FROM THE DENTAL LIGHT HAS A THREE PRONG LOCKING MECHANISM WHICH LOCKS THE LENS HEAT SHIELD/COVER INTO PLACE. THE LENS HEAT SHIELD/COVER HAS TO BE REMOVED BY THE END USER WHEN CLEANING THE LENS OR REPLACING THE HALOGEN BULB DURING ROUTINE OPERATOR MAINTENANCE. THE INSTALLATION INSTRUCTIONS, USE AND CARE INSTRUCTIONS, AND LABELING CLEARLY STATE TO ENSURE THE LENS HEAT SHIELD/HOLDER IS PROPERLY SECURE BY SNAPPING THE PART BACK INTO PLACE. DENTAL EQUIPMENT LLC (B)(4) INITIATED A RECALL ON JUNE 10, 2011 TO ADD A TETHER TO THE LENS HEAT SHIELD/COVER TO KEEP IT FROM FALLING OFF THE DENTAL LIGHT IF IT IS NOT RE-INSTALLED PROPERLY. THE FDA CLOSED THIS RECALL ON 10/26/2012 (PLEASE REFERENCE FDA RECALL Z-2834-2011). MARUS NOTIFIED THE DISTRIBUTOR OF THE RECALL NUMEROUS TIMES WHILE THE RECALL WAS OPEN. THE DISTRIBUTOR RESPONDED BACK TO MARUS AND ACKNOWLEDGED THE RECALL BUT APPARENTLY THE DISTRIBUTOR DID NOT INSTALL THE TETHER KIT AT THIS ONE PARTICULAR DENTAL PRACTICE. AFTER FOLLOWING BACK UP WITH THE DISTRIBUTOR, THE DISTRIBUTOR WENT TO THE DENTAL OFFICE AND INSTALLED THE TETHER KIT ON (B)(6) 2013.

Description of Event or Problem · 1

A DOCTOR WAS PERFORMING ROUTINE MEDICAL TREATMENT TO A PATIENT WHEN THE LENS HEAT SHIELD/HOLDER FROM HIS DENTAL LIGHT CAME OFF THE LIGHT AND FELL ONTO THE PATIENTS CHEST THEN ROLLED UP TO THEIR NECK. THE DENTIST GAVE THE PATIENT SOME OVER THE COUNTER NEOSPORIN TO PUT ON THEIR NECK WHERE THE LENS HEAT SHIELD/HOLDER TOUCHED. THERE WERE NO SERIOUS INJURIES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
23712 MAXSTAR UNIT MOUNT LIGHT DENTAL OPERATING LIGHT EAZ MARUS ML1000

Patients

Seq Age Sex Outcome Treatment
1 57 YR