MAXSTAR UNIT MOUNT LIGHT
Report
- Report Number
- 1017522-2013-00002
- Event Type
- Malfunction
- Date Received
- January 16, 2013
- Date of Event
- December 20, 2012
- Report Date
- December 21, 2012
- Manufacturer
- MARUS
- Product Code
- EAZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- DENTIST
Narratives
THE LENS HEAT SHIELD/HOLDER FROM THE DENTAL LIGHT HAS A THREE PRONG LOCKING MECHANISM WHICH LOCKS THE LENS HEAT SHIELD/COVER INTO PLACE. THE LENS HEAT SHIELD/COVER HAS TO BE REMOVED BY THE END USER WHEN CLEANING THE LENS OR REPLACING THE HALOGEN BULB DURING ROUTINE OPERATOR MAINTENANCE. THE INSTALLATION INSTRUCTIONS, USE AND CARE INSTRUCTIONS, AND LABELING CLEARLY STATE TO ENSURE THE LENS HEAT SHIELD/HOLDER IS PROPERLY SECURE BY SNAPPING THE PART BACK INTO PLACE. DENTAL EQUIPMENT LLC (B)(4) INITIATED A RECALL ON JUNE 10, 2011 TO ADD A TETHER TO THE LENS HEAT SHIELD/COVER TO KEEP IT FROM FALLING OFF THE DENTAL LIGHT IF IT IS NOT RE-INSTALLED PROPERLY. THE FDA CLOSED THIS RECALL ON 10/26/2012 (PLEASE REFERENCE FDA RECALL Z-2834-2011). MARUS NOTIFIED THE DISTRIBUTOR OF THE RECALL NUMEROUS TIMES WHILE THE RECALL WAS OPEN. THE DISTRIBUTOR RESPONDED BACK TO MARUS AND ACKNOWLEDGED THE RECALL BUT APPARENTLY THE DISTRIBUTOR DID NOT INSTALL THE TETHER KIT AT THIS ONE PARTICULAR DENTAL PRACTICE. AFTER FOLLOWING BACK UP WITH THE DISTRIBUTOR, THE DISTRIBUTOR WENT TO THE DENTAL OFFICE AND INSTALLED THE TETHER KIT ON (B)(6) 2013.
A DOCTOR WAS PERFORMING ROUTINE MEDICAL TREATMENT TO A PATIENT WHEN THE LENS HEAT SHIELD/HOLDER FROM HIS DENTAL LIGHT CAME OFF THE LIGHT AND FELL ONTO THE PATIENTS CHEST THEN ROLLED UP TO THEIR NECK. THE DENTIST GAVE THE PATIENT SOME OVER THE COUNTER NEOSPORIN TO PUT ON THEIR NECK WHERE THE LENS HEAT SHIELD/HOLDER TOUCHED. THERE WERE NO SERIOUS INJURIES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 23712 | MAXSTAR UNIT MOUNT LIGHT | DENTAL OPERATING LIGHT | EAZ | MARUS | ML1000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR |