FDA Adverse Event
Malfunction
Summary report: N
SECURE II MED/SURG BED
MDR report key: 2926291
·
Received January 23, 2013
Report
- Report Number
- 0001831750-2013-00309
- Event Type
- Malfunction
- Date Received
- January 23, 2013
- Date of Event
- January 3, 2013
- Report Date
- January 3, 2013
- Manufacturer
- STRYKER MEDICAL-KALAMAZOO
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
RESULT - LIFT CABLE CONNECTORS PLUGGED IN BACKWARDS ON CPU BOARD.
Description of Event or Problem · 1
IT WAS REPORTED BY SERVICE REPORT THAT TREND MOTION WAS REVERSED WHEN PRESSING BUTTON AND BED LOSES ABILITY TO LEVEL OUT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 31789 | SECURE II MED/SURG BED | BED, AC-POWERED ADJUSTABLE HOSPITAL | FNL | STRYKER MEDICAL-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |