FDA Adverse Event Injury Summary report: N

SYNFIX MINI-OPEN IMPLANT COUPLING

MDR report key: 2926047 · Received January 23, 2013

Report

Report Number
1719045-2013-00177
Event Type
Injury
Date Received
January 23, 2013
Date of Event
January 2, 2013
Report Date
January 2, 2013
Manufacturer
SYNTHES MONUMENT
Product Code
MAX
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED.

Additional Manufacturer Narrative · 1

DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. NEMCOMED, PRESENTLY AVALIGN NEMCOMED, MANUFACTURED THE SYNFIX MINI-OPEN IMPLANT COUPLING, THE COUPLINGS WERE INSPECTED AND CONFORMED TO THE SYNTHES INSPECTION SHEET. THERE WERE NO COMPLAINT-RELATED ANOMALIES. THE THREADED SHAFT OF THE COUPLING IS FRACTURED WITH APPROXIMATELY 11.5MM REMAINING IN THE IMPLANT. THE FRACTURED PORTION WHICH REMAINS IN THE IMPLANT HAS DAMAGED THREADS AND THE REMAINING SHAFT IS DAMAGED AS WELL. IT IS PROTRUDING OUT OF THE IMPLANT OFF-AXIS. THE OFF-AXIS ORIENTATION OF THE BROKEN THREAD SHAFT INDICATES THAT COUPLING WAS CROSS THREADED. THE CROSS THREADING LEAD TO THE OVER TIGHTENING OF THE THREADED SHAFT. THE OVER TIGHTENING LEAD TO TORSIONAL STRESSES EXCEEDING THE STRENGTH OF THE OF THE WEAKEST POINT ON THE SHAFT WHICH IS 2MM IN DIAMETER. THE DAMAGED THREADS ARE MOST LIKELY DUE TO THE REMOVAL OF THE IMPLANT USING OR INSTRUMENTS.

Description of Event or Problem · 1

AS THE SURGEON WAS TRYING TO ATTACH THE MINI OPEN GUIDE DURING AN ALIF FUSION L5 TO S1 WITH SYNFIX LR, THE INNER SPINDLE SNAPPED OFF IN THE IMPLANT AND REMAINED STUCK TO IT. THE ENTIRE IMPLANT WAS PULLED OUT AND A NEW IMPLANT INSERTED. SURGERY WAS PROLONGED 10 TO 15 MINUTES. NO ADVERSE EFFECT TO THE PATIENT WAS REPORTED. THIS IS 2 OF 2 REPORTS FOR THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
33060 SYNFIX MINI-OPEN IMPLANT COUPLING IMPLANT COUPLING MAX SYNTHES MONUMENT 80252

Patients

Seq Age Sex Outcome Treatment
1 47 YR Required Intervention SYNFIX-LR, IMPLANT COUPLING