SYNFIX MINI-OPEN IMPLANT COUPLING
Report
- Report Number
- 1719045-2013-00177
- Event Type
- Injury
- Date Received
- January 23, 2013
- Date of Event
- January 2, 2013
- Report Date
- January 2, 2013
- Manufacturer
- SYNTHES MONUMENT
- Product Code
- MAX
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED.
DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. NEMCOMED, PRESENTLY AVALIGN NEMCOMED, MANUFACTURED THE SYNFIX MINI-OPEN IMPLANT COUPLING, THE COUPLINGS WERE INSPECTED AND CONFORMED TO THE SYNTHES INSPECTION SHEET. THERE WERE NO COMPLAINT-RELATED ANOMALIES. THE THREADED SHAFT OF THE COUPLING IS FRACTURED WITH APPROXIMATELY 11.5MM REMAINING IN THE IMPLANT. THE FRACTURED PORTION WHICH REMAINS IN THE IMPLANT HAS DAMAGED THREADS AND THE REMAINING SHAFT IS DAMAGED AS WELL. IT IS PROTRUDING OUT OF THE IMPLANT OFF-AXIS. THE OFF-AXIS ORIENTATION OF THE BROKEN THREAD SHAFT INDICATES THAT COUPLING WAS CROSS THREADED. THE CROSS THREADING LEAD TO THE OVER TIGHTENING OF THE THREADED SHAFT. THE OVER TIGHTENING LEAD TO TORSIONAL STRESSES EXCEEDING THE STRENGTH OF THE OF THE WEAKEST POINT ON THE SHAFT WHICH IS 2MM IN DIAMETER. THE DAMAGED THREADS ARE MOST LIKELY DUE TO THE REMOVAL OF THE IMPLANT USING OR INSTRUMENTS.
AS THE SURGEON WAS TRYING TO ATTACH THE MINI OPEN GUIDE DURING AN ALIF FUSION L5 TO S1 WITH SYNFIX LR, THE INNER SPINDLE SNAPPED OFF IN THE IMPLANT AND REMAINED STUCK TO IT. THE ENTIRE IMPLANT WAS PULLED OUT AND A NEW IMPLANT INSERTED. SURGERY WAS PROLONGED 10 TO 15 MINUTES. NO ADVERSE EFFECT TO THE PATIENT WAS REPORTED. THIS IS 2 OF 2 REPORTS FOR THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 33060 | SYNFIX MINI-OPEN IMPLANT COUPLING | IMPLANT COUPLING | MAX | SYNTHES MONUMENT | 80252 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | Required Intervention | SYNFIX-LR, IMPLANT COUPLING |