OT VERIO TEST STRIPS
Report
- Report Number
- 3008382007-2013-01336
- Event Type
- Malfunction
- Date Received
- January 23, 2013
- Report Date
- December 28, 2012
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4): THE TEST STRIP PASSED TESTING WITH NO FAULTS FOUND.IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
THIS CORRECTION IS BEING SENT TO ADD ADDITIONAL INFORMATION THAT WAS MISSING FROM A PREVIOUS SUPPLEMENTAL REPORT.FOLLOW-UP # 3 (6/25/2013)-DEVICE EVALUATION: THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 5/24/2013 WITH THE FOLLOWING FINDINGS: THE METER HAS PASSED TESTING. THE COMPLAINT RELATING TO THIS DEVICE COULD NOT BE REPRODUCED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(4) 2012, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING INACCURATE RESULTS COMPARED TO THE SAME METER WITH RESULTS OF "18.3, 12.2 AND 9.6MMOL/L." THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED RESULTS DID NOT MEET LIFESCAN'S ACCURACY/PRECISION CRITERIA AND THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 32275 | OT VERIO TEST STRIPS | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3386295 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |