FDA Adverse Event Malfunction Summary report: N

LARGE NEEDLE DRIVER INSTRUMENT

MDR report key: 2925921 · Received January 22, 2013

Report

Report Number
2955842-2013-00265
Event Type
Malfunction
Date Received
January 22, 2013
Date of Event
December 20, 2012
Report Date
December 26, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING WAS UNABLE TO CONFIRM THE CUSTOMER REPORTED FAILURE MODE AS THE DRIVE CABLES ARE NOT BROKEN OR FRAYED AT THE WRIST. THE GRIPS ARE FOUND TO OPEN/CLOSE PROPERLY WHEN THE INPUT DISCS ARE MANUALLY ROTATED. ENGINEERING EVALUATION ALSO FOUND THAT THE DISTAL END OF THE MAIN TUBE HAS VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES ARE .080 - .200 IN LENGTH AND NOT ALIGNED WITH THE TUBE AXIS. ENGINEERING CONCLUDED THAT THE DAMAGE IS LIKELY DUE TO MISHANDLING/MISUSE. NO OTHER DAMAGE FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI HYSTERECTOMY PROCEDURE, THE CABLE ON THE LARGE NEEDLE DRIVER INSTRUMENT BROKE. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
30398 LARGE NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420006-06 M10121015 754

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SURGICAL SYSTEM AND ACCESSORIES.