FDA Adverse Event Malfunction Summary report: N

COULTER® LH 750 HEMATOLOGY ANALYZER

MDR report key: 2925916 · Received January 22, 2013

Report

Report Number
1061932-2013-00111
Event Type
Malfunction
Date Received
January 22, 2013
Date of Event
December 28, 2012
Report Date
December 28, 2012
Manufacturer
BECKMAN COULTER, INC.
Product Code
GKZ
PMA / PMN Number
K011342
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) OBSERVED FLUID LEAKED IN THE HEMOGLOBIN (HGB) BLANK LINE AND REPLACED THE FITTING TO RESOLVE THE FLUID LEAK ISSUE. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED APPROXIMATELY THREE HUNDRED (300) MILLILITERS OF CLENZ CLEANER SOLUTION LEAKED FROM THE INSTRUMENT AND ONTO THE COUNTER DURING SYSTEM STARTUP INVOLVING THE COULTER LH 750 HEMATOLOGY ANALYZER. THE CUSTOMER HAD ON PROPER PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND EYE PROTECTION AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. PATIENT SAMPLES WERE NOT ANALYZED AT THE TIME OF THE EVENT. THERE WAS NO PATIENT IMPACT. THE CUSTOMER HAS AN EXPOSURE CONTROL/RISK MANAGEMENT PLAN AT THE FACILITY. BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
29735 COULTER® LH 750 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1