FDA Adverse Event
Malfunction
Summary report: N
PK DISSECTING FORCEPS INSTRUMENT
MDR report key: 2925903
·
Received January 22, 2013
Report
- Report Number
- 2955842-2013-00264
- Event Type
- Malfunction
- Date Received
- January 22, 2013
- Date of Event
- December 26, 2012
- Report Date
- December 26, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K061260
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE INSTRUMENT WAS RETURNED WITH A FRAYED PITCH CABLE AT DISTAL CLEVIS HUB. NO EXCESSIVE DAMAGE WAS FOUND AT THE CLEVIS. THE FRAYED SEGMENT IS APPROXIMATELY 0.03 IN LENGTH. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING REPROCESSING OF THE PK DISSECTING FORCEPS INSTRUMENT IT WAS NOTED THAT THE WIRE ON THE INSTRUMENT WAS BROKEN. THERE WERE NO MISSING OR FALLEN PIECES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 30323 | PK DISSECTING FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420227-03 | M10120831 353 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SURG SYSTEM, ACCESSORIES AND ESU |