FDA Adverse Event Malfunction Summary report: N

COULTER® LH 500 HEMATOLOGY ANALYZER

MDR report key: 2925900 · Received January 22, 2013

Report

Report Number
1061932-2013-00110
Event Type
Malfunction
Date Received
January 22, 2013
Date of Event
December 27, 2012
Report Date
December 27, 2012
Manufacturer
BECKMAN COULTER, INC.
Product Code
GKZ
PMA / PMN Number
K042724
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) OBSERVED A SMALL FLUID LEAK AROUND THE WHITE BLOOD CELL (WBC) HOUSING ASSEMBLY. THE FSE REPLACED THE WBC HOUSING AND WBC APERTURE TO RESOLVE THE ISSUE. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED APPROXIMATELY TEN (10) MILLILITERS OF BLUISH CLEAR FLUID LEAKED FROM THE INSTRUMENT INVOLVING THE COULTER LH 500 HEMATOLOGY ANALYZER. THE CUSTOMER HAD ON PROPER PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND EYE PROTECTION AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. PATIENT RESULTS WERE NOT IMPACTED AND NO ERRONEOUS RESULTS WERE GENERATED. BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
30322 COULTER® LH 500 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1