M-1 COT - BASE HIGH LOAD 28
Report
- Report Number
- 0001831750-2013-00273
- Event Type
- Malfunction
- Date Received
- January 22, 2013
- Date of Event
- October 20, 2012
- Report Date
- December 18, 2012
- Manufacturer
- STRYKER MEDICAL-KALAMAZOO
- Product Code
- FPO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DURING THE COURSE OF THE INVESTIGATION IT WAS NOTED THAT THE RETAINING POST WAS WORN. THE RETAINING POST (PN: (B)(4)) ON THE UNIT WAS REPLACED HOWEVER THIS WAS DETERMINED TO NOT HAVE CAUSED OR CONTRIBUTED TO THE EVENT. BASED ON THE EVENT DESCRIPTION, THE FOOT END LEGS DID NOT HOLD AND COLLAPSED WHEN THE COT WAS UNLOADED FROM THE AMBULANCE. THE UNIT WAS DESCRIBED TO BE TRANSPORTING AN INCUBATOR AT THE TIME THE EVENT OCCURRED. BASED ON A REVIEW OF PREVIOUS SIMILAR INVESTIGATIONS, A LIKELY CAUSE AS TO WHY THE COT COLLAPSED IS DUE TO THE LEGS NOT BEING VERIFIED TO BE LOCKED. THE MANUAL STATES TO ENSURE THAT THE LEGS ARE LOCKED DURING THE UNLOADING PROCESS. AN ADDITIONAL CONTRIBUTING FACTOR TO THE COT DROP EVENT IS DUE TO THE UNIT BEING USED TO TRANSPORT AN INCUBATOR. THE M1 COT IS NOT DESIGNED TO BE USED TO TRANSPORT AN INCUBATOR. ANOTHER POSSIBLE CONTRIBUTING FACTOR TO THE DROP EVENT MAY BE DUE TO THE COT BEING USED PAST ITS LIFE EXPECTANCY.
OUR CUSTOMER USED THE M1 COT WITH A TRANSPORT INCUBATOR (DRAGERTI500). DURING UNLOADING FROM THE AMBULANCE THE DEVICE COMBINATION OVERTURNED, BECAUSE THE COT COLLAPSED AT THE FOOT END. THE INCUBATOR WAS TURNED UP AND THE TRANSPORTATION WAS CONTINUED BY HAND.
OUR CUSTOMER USED THE M1 COT WITH A TRANSPORT INCUBATOR (DRÄGERTI500). DURING UNLOADING FROM THE AMBULANCE THE DEVICE COMBINATION OVERTURNED, BECAUSE THE COT COLLAPSED AT THE FOOT END. THE INCUBATOR WAS TURNED UP AND THE TRANSPORTATION WAS CONTINUED BY HAND.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 30197 | M-1 COT - BASE HIGH LOAD 28 | STRETCHER, WHEELED | FPO | STRYKER MEDICAL-KALAMAZOO | 6100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |