FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2925593 · Received January 22, 2013

Report

Report Number
2531779-2013-01039
Event Type
Malfunction
Date Received
January 22, 2013
Date of Event
January 2, 2013
Report Date
January 10, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Additional Manufacturer Narrative · 1

(B)(4) - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD COVER WAS RETURNED PEELING AT THE OK BUTTON. A DAMAGED KEYPAD WILL PERMIT CONTAMINATION TO PERMEATE THE BUTTONS WHICH WILL HAVE A NEGATIVE IMPACT ON BUTTON FUNCTION. THIS SITUATION IS NOT LIKELY TO RESULT IN AN ADVERSE EVENT AS THE DAMAGED KEYPAD SHOULD BE CLEARLY VISIBLE AND WARNS THE PATIENT TO DISCONTINUE USING THE PUMP. DURING TESTING, THE OK, UP ARROW, AND DOWN ARROW KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. DURING INVESTIGATION, EVIDENCE OF CONTAMINATION WAS FOUND UNDER ALL KEY CONTACTS. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED A DISCOLORED/FADED DISPLAY SCREEN, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION.

Description of Event or Problem · 1

ON (B)(4) 2013, THE DISTRIBUTOR CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
29570 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1